dc.identifier.citationLell, J. A., 2023: Desarrollo de software y hardware para la puesta en valor de adquisidores de datos Campbell Scientific CR10X. Nota Técnica SMN 2023-141.es
dc.identifier.urihttp://hdl.handle.net/20.500.12160/2478
dc.descriptionFil: Lell, Julián. Servicio Meteorológico Nacional. Dirección Nacional de Ciencia e Innovación en Productos y Servicios. Dirección Central de Monitoreo del Clima; Argentina.es
dc.description.abstractLos adquisidores de datos Campbell Scientific Inc. modelo CR10X fueron introducidos al mercado en 1995, y discontinuados probablemente en 2004, con la introducción del modelo CR1000. Desde un punto de vista instrumental, los adquisidores CR10X siguen siendo equipos de gran calidad y precisión (estando debidamente calibrados), pero han quedado obsoletos desde un punto de vista operativo: el método de programación requiere comandos ad hoc, que podrían considerarse “de bajo nivel”, no pudiendo utilizarse el lenguaje de alto nivel “CRBasic” (introducido en 2000 con el adquisidor modelo CR5000) de mayor versatilidad y utilizado hasta la actualidad. Adicionalmente, los dispositivos CR10X poseen un único puerto de conexión “CS I/O”, propietario de Campbell Scientific Inc., y carecen de otras opciones de conectividad como puertos RS-232, Ethernet, USB, etc.; esta carencia de alternativas de conectividad hace que la programación y recuperación de datos requiera de accesorios tales como el “SC32A” (interfaz RS-232 optoacoplada), el cual ha sido discontinuado por la firma. En el presente documento se presenta un desarrollo que permite el acceso completo al adquisidor CR10X a través de un programa en lenguaje Python, utilizando un adaptador implementado en un módulo “Arduino UNO”, de manera que resulta posible utilizar este modelo de adquisidores en cualquier tipo de plataforma (Linux, Windows, etc.), sin la necesidad de contar con accesorios o software original de la firma Campbell Sci. Inc..es
dc.description.abstractCampbell Scientific’s CR10X dataloggers were introduced on the market in 1995, and discontinued in 2004, with the introduction of CR1000 model. From an instrumental point of view, CR10X dataloggers are devices of great quality and accuracy even to this day (provided they’re properly calibrated), but they’ve been rendered obsolete from a functional point of view: the programming method requires ad hoc commands, which can be considered low-level, not being possible the use of the high-level language “CRBasic” (introduced in 2000 with the model CR5000), of greater versatility and used in the present day. On the other hand, CR10X devices feature a single connection port “CS I/O”, Campbell Sci. Inc. proprietary, and lack any other connectivity option like RS-232 ports, Ethernet, USB, etc.; this lack of connectivity alternatives makes the programming and retrieval of data to require accessories such as the “SC32A” (opto-coupled RS-232 interface), which has been discontinued by the company. This document presents a development which allows for complete access to CR10X datalogger through a script written in Python language, using a hardware adapter implemented in “Arduino UNO” module, making possible the use of this model of dataloggers with any kind of platform (Linux, Windows, etc.), without the need of accessories neither original software from Campbell Sci. Inc..en
dc.formatapplication/pdf
dc.language.isospaes
dc.publisherServicio Meteorologico Nacional. Dirección Nacional de Ciencia e Innovación en Productos y Servicios. Dirección Central de Monitoreo del Clima.es
dc.rightsinfo:eu-repo/semantics/openAccesses
dc.rightshttp://creativecommons.org/licenses/by/2.5/ar/es
dc.rightsinfo:eu-repo/semantics/openAccesses
dc.subjectCR10Xes
dc.subjectDATALOGGERes
dc.subjectCAMPBELL SCIENTIFICes
dc.subjectACTUALIZACIÓNes
dc.subjectPUESTA EN VALORes
dc.titlePuesta en valor de adquisidores de datos Campbell Scientific Inc. CR10Xes